Test Code ACHRP Acetylcholine Receptor Antibody Reflexive Panel

Clinical Indications

Preferred reflexive panel for diagnosing myasthenia gravis. Panel contains binding, blocking, and modulating antibodies.

Additional Test Codes

Epic Order Name: ACETYLCH REC PANL AB (ACHRP)
Sunquest Order Code: ACHRP
Epic Px Code: LAB2012

Specimen Collection Type

Gold tube
 

Minimum Collection Volume

1.5 mL

Reference Range

Components

Reference Interval

Acetylcholine Receptor Binding Antibody

Negative

0.0-0.4 nmol/L

Positive

0.5 nmol/L or greater

Acetylcholine Receptor Blocking Antibody

Effective November 18, 2013

Negative:

0-26% blocking

Indeterminate:

27-41% blocking

Positive:

42% or greater blocking

Acetylcholine Receptor Modulating Antibody

Effective August 20, 2012

Negative

0-45% modulating

Positive

46% or greater modulating

 

Critical Value

n/a

Interpretation

Approximately 85-90 percent of patients with myasthenia gravis (MG) express antibodies to the acetylcholine receptor (AChR), which can be divided into binding, blocking, and modulating antibodies. Binding antibody can activate complement and lead to loss of AChR. Blocking antibody may impair binding of acetylcholine to the receptor, leading to poor muscle contraction. Modulating antibody causes receptor endocytosis resulting in loss of AChR expression, which correlates most closely with clinical severity of disease. Approximately 10-15 percent of individuals with confirmed myasthenia gravis have no measurable binding, blocking, or modulating clinical diagnosis or patient management decisions.

Rejection Criteria

Contaminated, hemolyzed, or severely lipemic specimens

Specimen Availability for Add-on Testing

2 Days

Specimen Stability

Refrigerated (preferred): 2 weeks
Ambient: 2 hours
Frozen: 1 year

Day(s) Performed

Monday through Sunday

Performing Lab - Department

TUKHS-BH-Mail Outs

CPT Code

Acetylcholine Receptor Binding Antibody: 83519
Acetylcholine Receptor Blocking Antibody: 83516
Acetylcholine Receptor Modulating Antibody (if reflexed): add 83516

LOINC Code

Acetylcholine Receptor Binding Antibody: 11034-6
Acetylcholine Receptor Blocking Antibody: 42937-3
Acetylcholine Receptor Modulating Antibody (if reflexed): add 30192-9

Test Alias

AChR Antibody
AChR reflex panel
Muscle nicotinic Acetylcholine Receptor (AChR) Binding Antibody
Myasthenia Gravis Antibodies
Myasthenia Gravis Antibodies Reflex

Additional Information / Notes

If Acetylcholine Receptor Binding Antibody result is greater than 0.4 nmol/L or Acetylcholine Receptor Blocking Antibody result is greater than 26 percent, then Acetylcholine Receptor Modulating Antibody (ARUP test code 0099521) will be added. Additional charges apply.