Test Code SYPT Treponema pallidum (Syphilis) Total Antibodies
Additional Test Codes
Epic Order Name: Syphilis Ab Screen
Sunquest Order Code: SYPT
Epic Px Code: LAB2719
Specimen Collection Type
Preferred: Gold
Acceptable: Red
Minimum Collection Volume
0.5 mL Serum
Reference Range
Negative
Rejection Criteria
Gross hemolysis
Specimen Availability for Add-on Testing
36 Hours
Specimen Stability
Room temperature: 5 days
Refrigerated: 14 days
Frozen: 12 months
Day(s) Performed
Monday-Sunday 8am-4pm
Performing Lab- Department
TUKHS-BH-CHEMISTRY
CPT Code
86780
LOINC Code
47236-5
Test Alias
Syphilis Serology
Treponema pallidum (T. pallidum)
Additional Information / Notes
Positive result Auto-reflexes to RPR
Clinical Indications
N/A
Critical Value
N/A
Interpretation
Syphilis Serology Interpretation Guidelines |
|||
EIA Screen (Treponemal) |
RPR (Non-Treponemal) |
TPPA (Treponemal) |
Interpretation |
Nonreactive |
N/A |
N/A |
Negative for syphilis. No further testing required, unless clinically indicated. |
Reactive |
Reactive |
N/A |
Consistent with untreated or recently treated syphilis. See CDC treatment guidelines.
Follow up RPR titers: Order RPRT |
Reactive |
Nonreactive |
Reactive |
Consistent with past successfully treated syphilis. Early or latent syphilis possible in individuals with no history of treated syphilis. |
Reactive |
Nonreactive |
Nonreactive |
Consistent with false positive screening test. No further testing required, unless clinically indicated.
|